A high - performance liquid chromatographic method for the determination of alfuzosin in human plasma has been developed and validated. A column-switching procedure without extraction was used to isolate the drug from biological matrix prior to the quantitative analysis. The lower limit of detection for the analyte was 1 ng/ml. The method was linear from 2 to 150 ng/ml for human plasma. Within- and between-assay precision and accuracy were all found to be <5.2% at the eight concentrations evaluated. This procedure, simple and rapid, is suitable for pharmacology studies on alfuzosin.
Determination of Alfuzosin in Human Plasma by High-Performance Liquid Chromatography with Column-Switching
CARLUCCI, Giuseppe;MAZZEO, PIETRO
1994-01-01
Abstract
A high - performance liquid chromatographic method for the determination of alfuzosin in human plasma has been developed and validated. A column-switching procedure without extraction was used to isolate the drug from biological matrix prior to the quantitative analysis. The lower limit of detection for the analyte was 1 ng/ml. The method was linear from 2 to 150 ng/ml for human plasma. Within- and between-assay precision and accuracy were all found to be <5.2% at the eight concentrations evaluated. This procedure, simple and rapid, is suitable for pharmacology studies on alfuzosin.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.