PURPOSE: The aim of this study was to assess whether there was a decrease of bacterial leakage with increasing torque values in conical Morse Cone connection implants. METHODS: A total of 30 Morse Cone conical tapered implants (10 implants per group) were used in this study. The abutments were connected to the implants with 20 N (group 1), 30 N (group 2), and 40 N (group 3) insertion torque values. The inner parts of 5 implants, per group, were inoculated with Pseudomonas aeruginosa suspension and the remaining 5 implants, per group, with Aggregatibacter actinomycetemcomitans. The penetration of bacteria into the surrounding solution was determined by the observation of turbidity of the broth. RESULTS: In groups 1 and 2, bacterial contamination was found in 2 of the 10 implants, only in the specimens seeded with P. aeruginosa. In group 3, no contaminated samples were found.

Bacterial Leakage in Morse Cone Internal Connection Implants Using Different Torque Values: An In Vitro Study

D'ERCOLE, Simonetta;TRIPODI, Domenico;PERROTTI, Vittoria;PIATTELLI, Adriano;IEZZI, GIOVANNA
2014-01-01

Abstract

PURPOSE: The aim of this study was to assess whether there was a decrease of bacterial leakage with increasing torque values in conical Morse Cone connection implants. METHODS: A total of 30 Morse Cone conical tapered implants (10 implants per group) were used in this study. The abutments were connected to the implants with 20 N (group 1), 30 N (group 2), and 40 N (group 3) insertion torque values. The inner parts of 5 implants, per group, were inoculated with Pseudomonas aeruginosa suspension and the remaining 5 implants, per group, with Aggregatibacter actinomycetemcomitans. The penetration of bacteria into the surrounding solution was determined by the observation of turbidity of the broth. RESULTS: In groups 1 and 2, bacterial contamination was found in 2 of the 10 implants, only in the specimens seeded with P. aeruginosa. In group 3, no contaminated samples were found.
File in questo prodotto:
File Dimensione Formato  
2014 D'Ercole-implant dent.pdf

Solo gestori archivio

Tipologia: Documento in Post-print
Dimensione 218.81 kB
Formato Adobe PDF
218.81 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11564/650533
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 26
  • ???jsp.display-item.citation.isi??? 22