PURPOSE: To investigate whether the interposition of a sealing-connector was able to reduce the bacterial leakage in external hexagon implants. METHODS: 20 implants with external hexagon connection were used. Ten Test implant-abutment assemblies were connected with the interposition of a sealing-connector molded in the exact shape of the two opposed surfaces. Ten Control implant-abutment assemblies were connected with no sealing-connector interposed. Two types of bacteria were introduced into the internal portion of the implant, before placing the connector. The study lasted 28 days. RESULTS: All control specimens, seeded with P. aeruginosa (PA) and A. actynomycetemcomitans (AA), showed contamination of the culture medium, indicative of microbial leakage. In the Test specimens, three instances of contaminated specimens were found in the samples seeded with PA and two contaminated specimens in the ones seeded with AA, for a total of five contaminated samples out of 10. The use of the sealing-connector was able to prevent bacterial leakage in half of the samples (50%). The leakage in both groups occurred mainly in the last week of the experiment. Probably, a longer period, under the conditions of this experiment, is necessary for the migration of the bacteria, and, furthermore, an observation period of 7 or 14 days may not be enough to show microbial contamination.
Bacterial leakage in external hexagon implants with and without the interposition of a sealing-connector
DOMENICO TRIPODI;FLAVIA IACULLI;ADRIANO PIATTELLI;GIOVANNA IEZZI;VITTORIA PERROTTI
Penultimo
;SIMONETTA D’ERCOLEUltimo
2018-01-01
Abstract
PURPOSE: To investigate whether the interposition of a sealing-connector was able to reduce the bacterial leakage in external hexagon implants. METHODS: 20 implants with external hexagon connection were used. Ten Test implant-abutment assemblies were connected with the interposition of a sealing-connector molded in the exact shape of the two opposed surfaces. Ten Control implant-abutment assemblies were connected with no sealing-connector interposed. Two types of bacteria were introduced into the internal portion of the implant, before placing the connector. The study lasted 28 days. RESULTS: All control specimens, seeded with P. aeruginosa (PA) and A. actynomycetemcomitans (AA), showed contamination of the culture medium, indicative of microbial leakage. In the Test specimens, three instances of contaminated specimens were found in the samples seeded with PA and two contaminated specimens in the ones seeded with AA, for a total of five contaminated samples out of 10. The use of the sealing-connector was able to prevent bacterial leakage in half of the samples (50%). The leakage in both groups occurred mainly in the last week of the experiment. Probably, a longer period, under the conditions of this experiment, is necessary for the migration of the bacteria, and, furthermore, an observation period of 7 or 14 days may not be enough to show microbial contamination.File | Dimensione | Formato | |
---|---|---|---|
Rizzo AJD Oct 2018.pdf
Solo gestori archivio
Tipologia:
PDF editoriale
Dimensione
343.01 kB
Formato
Adobe PDF
|
343.01 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.