The loosening of an abutment screw is one of the most frequent complications in implant-prosthetic rehabilitation, especially for single-crown cemented prostheses. This complication is due to several mechanical factors including type of connection, abutment-screw geometry, settling effects, and cyclical load. The purpose of the present in vitro study was to compare and associate different times of retightening with reductions in preload losses. We evaluated 40 internal hexagon dental implants and 40 external hexagon dental implants, with their related abutment screws. The implants were embedded in acrylic resin in cylindrical polyvinyl chloride tubes (26 mm diameter, 20 mm height). The abutments were fixed to the implants with screws to an initial torque of 35 Ncm using a digital torque meter with decimal precision. Two different types of connection were randomly divided in 4 subgroups of 10 samples each. One subgroup was used as control. The test groups underwent retightening to the same initial torque at increasing times from initial torque application for tightening of the abutment screws, to their retightening at 2 minutes, 5 minutes, and 10 minutes. The retightening time of 2 minutes shows significantly reduced preload loss. Randomized clinical trials are strongly required to provide clinicians with a beneficial standardized protocol of retightening that can be applied in routine clinical practice.
The relationship between time of re-tightening and preload loss of abutment screws for two different implant designs: an in-vitro study
Varvara, Giuseppe
;Sinjari, Bruna;Caputi, Sergio;Scarano, Antonio;Piattelli, Maurizio
2020-01-01
Abstract
The loosening of an abutment screw is one of the most frequent complications in implant-prosthetic rehabilitation, especially for single-crown cemented prostheses. This complication is due to several mechanical factors including type of connection, abutment-screw geometry, settling effects, and cyclical load. The purpose of the present in vitro study was to compare and associate different times of retightening with reductions in preload losses. We evaluated 40 internal hexagon dental implants and 40 external hexagon dental implants, with their related abutment screws. The implants were embedded in acrylic resin in cylindrical polyvinyl chloride tubes (26 mm diameter, 20 mm height). The abutments were fixed to the implants with screws to an initial torque of 35 Ncm using a digital torque meter with decimal precision. Two different types of connection were randomly divided in 4 subgroups of 10 samples each. One subgroup was used as control. The test groups underwent retightening to the same initial torque at increasing times from initial torque application for tightening of the abutment screws, to their retightening at 2 minutes, 5 minutes, and 10 minutes. The retightening time of 2 minutes shows significantly reduced preload loss. Randomized clinical trials are strongly required to provide clinicians with a beneficial standardized protocol of retightening that can be applied in routine clinical practice.File | Dimensione | Formato | |
---|---|---|---|
The relationship between time.pdf
Solo gestori archivio
Descrizione: Research Article
Tipologia:
Documento in Pre-print
Dimensione
1.85 MB
Formato
Adobe PDF
|
1.85 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.