Continuous flow left ventricular assist devices (LVADs) have become a valuable therapy for end-stage heart failure. In vitro research highlighted a role of outflow cannula position on the pattern of blood flow in the aorta. However, the clinical effects of the alterations of flow remain unclear. We investigate short- and long-term outcomes of patients implanted with Jarvik 2000 LVAD, according to the ascending (Group 1) versus descending (Group 2) outflow graft connection to the aorta in a multicenter study. From May 2008 to October 2014, 140 consecutive end-stage heart failure patients underwent Jarvik 2000 LVAD implantation in 17 Italian centers. According with a preliminary multivariate analysis, we selected the 90 patients implanted in the four high-volume centers to avoid bias (Group 1 n = 39, Group 2 n = 51). Among the groups, no differences were recorded in the hospital mortality and the main complications occurring after LVAD implantation were similar. In multivariable analysis, the ascending aorta outflow cannula position and higher creatinine at discharge were significant predictors for long-term survival. Postimplant hemolysis was more pronounced in descending aorta outflow graft anastomosis. Outflow graft anastomosis to the ascending aorta is associated with better long-term survival, independent of age and perfusion techniques, reflecting the previous in vitro results.
From bench to bedside: Impact of left ventricular assist device outflow conduit anastomosis position on outcome
Gabriele Di GiammarcoSecondo
;
2021-01-01
Abstract
Continuous flow left ventricular assist devices (LVADs) have become a valuable therapy for end-stage heart failure. In vitro research highlighted a role of outflow cannula position on the pattern of blood flow in the aorta. However, the clinical effects of the alterations of flow remain unclear. We investigate short- and long-term outcomes of patients implanted with Jarvik 2000 LVAD, according to the ascending (Group 1) versus descending (Group 2) outflow graft connection to the aorta in a multicenter study. From May 2008 to October 2014, 140 consecutive end-stage heart failure patients underwent Jarvik 2000 LVAD implantation in 17 Italian centers. According with a preliminary multivariate analysis, we selected the 90 patients implanted in the four high-volume centers to avoid bias (Group 1 n = 39, Group 2 n = 51). Among the groups, no differences were recorded in the hospital mortality and the main complications occurring after LVAD implantation were similar. In multivariable analysis, the ascending aorta outflow cannula position and higher creatinine at discharge were significant predictors for long-term survival. Postimplant hemolysis was more pronounced in descending aorta outflow graft anastomosis. Outflow graft anastomosis to the ascending aorta is associated with better long-term survival, independent of age and perfusion techniques, reflecting the previous in vitro results.File | Dimensione | Formato | |
---|---|---|---|
aor.13809.pdf
Solo gestori archivio
Descrizione: Articolo principale accettato per la pubblicazione
Tipologia:
Documento in Post-print
Dimensione
423.33 kB
Formato
Adobe PDF
|
423.33 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Artificial Organs - 2020 - Tarzia - From bench to bedside Impact of left ventricular assist device outflow conduit.pdf
Solo gestori archivio
Tipologia:
PDF editoriale
Dimensione
638.06 kB
Formato
Adobe PDF
|
638.06 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.