Background: Posterior maxillary atrophies could emerge after the loss of teeth, trauma, infections, or lesions that often require regenerative approaches. In these critical conditions, the achievement of implant primary stability represents a clinical challenge in the operative practice. Therefore, a two-stage approach is often preferred with a delay of the rehabilitation time and a consistent increasing of the biological and the operative costs. The aim of this study was to evaluate the mechanical behaviour of a self-condenser implant compared to a standard implant in a critical simulation on different thicknesses and densities of polyurethane lamina. Materials and methods: A total of two implant models were tested: a self-condensing device (test) and a standard implant (control). The study evaluated the insertion torque and the pull-out strength values of the test and control implants inserted in different sizes (1, 2, and 3 mm) and density polyurethane lamina (10, 20, and 30 pcf) for a total of 320 experimental sites. Results: In total, 320 experimental sites were produced in the polyurethane samples. A statistically significant difference of insertion and pull-out torque values between the test and control Implants was found in the different bone densities (p < 0.05). The insertion and pull-out torque values were always higher for the test implants in all experimental conditions. In all bone densities, the insertion torque values were higher than the pull-out torque values. The self-condenser dental implant design evaluated in this in vitro study showed a high level of stability in all experimental conditions. Conclusions: The test implant could represent a useful tool for a one-stage surgical approach in the presence of limited residual native bone as an alternative to a delayed technique.
Mechanical Behaviour and Primary Stability of a Self-Condensing Implant: A Laboratory Critical Simulation of a Severe Maxillary Atrophy on Polyurethane Lamina
Iezzi G.;Di Pietro N.Co-primo
;D'arcangelo C.;Piattelli A.;Tumedei M.
2022-01-01
Abstract
Background: Posterior maxillary atrophies could emerge after the loss of teeth, trauma, infections, or lesions that often require regenerative approaches. In these critical conditions, the achievement of implant primary stability represents a clinical challenge in the operative practice. Therefore, a two-stage approach is often preferred with a delay of the rehabilitation time and a consistent increasing of the biological and the operative costs. The aim of this study was to evaluate the mechanical behaviour of a self-condenser implant compared to a standard implant in a critical simulation on different thicknesses and densities of polyurethane lamina. Materials and methods: A total of two implant models were tested: a self-condensing device (test) and a standard implant (control). The study evaluated the insertion torque and the pull-out strength values of the test and control implants inserted in different sizes (1, 2, and 3 mm) and density polyurethane lamina (10, 20, and 30 pcf) for a total of 320 experimental sites. Results: In total, 320 experimental sites were produced in the polyurethane samples. A statistically significant difference of insertion and pull-out torque values between the test and control Implants was found in the different bone densities (p < 0.05). The insertion and pull-out torque values were always higher for the test implants in all experimental conditions. In all bone densities, the insertion torque values were higher than the pull-out torque values. The self-condenser dental implant design evaluated in this in vitro study showed a high level of stability in all experimental conditions. Conclusions: The test implant could represent a useful tool for a one-stage surgical approach in the presence of limited residual native bone as an alternative to a delayed technique.File | Dimensione | Formato | |
---|---|---|---|
Comuzzi and Di Pietro et al Appl Sc 2022.pdf
accesso aperto
Descrizione: Article
Tipologia:
PDF editoriale
Dimensione
3.19 MB
Formato
Adobe PDF
|
3.19 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.