Background: The primary stability is a determinant clinical condition for the success of different dental implants macro-design in different bone density using a validated and repeatable in vitro technique employing solid rigid polyurethane blocks. Materials and Methods: Five implants 3.8 x 13 mm(2) for each macro-design (i.e., IK-tapered; IC-cylindric; and IA-active blade shape) were positioned into 20- and 30- pounds per cubic foot (PCF) polyurethane blocks. Bucco-lingual (BL) and mesial-distal (MD) implant stability quotient score (ISQ) was assessed by resonance frequency analysis while, insertion/removal torques were evaluated by dynamometric ratchet. Results: IC implants shown better primary stability in terms of ISQ compared to IA and IK in lower density block (20 PCF), while IK was superior to IA in higher density (30 PCF). IC shown higher removal torque in 30-PCF compared to IA and IC. Conclusions: The study effectiveness on polyurethane artificial bone with isotropic symmetry structure showed that the implants macro-design might represent a key factor on primary stability, in particular on low-density alveolar bone. Clinicians should consider patients features and implant geometry during low-density jaws rehabilitation. Further investigations are needed to generalize these findings.
The Influence of the Implant Macrogeometry on Insertion Torque, Removal Torque, and Periotest Implant Primary Stability: A Mechanical Simulation on High-Density Artificial Bone
Tumedei, Margherita
Primo
;Petrini, MorenaSecondo
;Piattelli, Adriano
Penultimo
;Iezzi, GiovannaUltimo
2021-01-01
Abstract
Background: The primary stability is a determinant clinical condition for the success of different dental implants macro-design in different bone density using a validated and repeatable in vitro technique employing solid rigid polyurethane blocks. Materials and Methods: Five implants 3.8 x 13 mm(2) for each macro-design (i.e., IK-tapered; IC-cylindric; and IA-active blade shape) were positioned into 20- and 30- pounds per cubic foot (PCF) polyurethane blocks. Bucco-lingual (BL) and mesial-distal (MD) implant stability quotient score (ISQ) was assessed by resonance frequency analysis while, insertion/removal torques were evaluated by dynamometric ratchet. Results: IC implants shown better primary stability in terms of ISQ compared to IA and IK in lower density block (20 PCF), while IK was superior to IA in higher density (30 PCF). IC shown higher removal torque in 30-PCF compared to IA and IC. Conclusions: The study effectiveness on polyurethane artificial bone with isotropic symmetry structure showed that the implants macro-design might represent a key factor on primary stability, in particular on low-density alveolar bone. Clinicians should consider patients features and implant geometry during low-density jaws rehabilitation. Further investigations are needed to generalize these findings.File | Dimensione | Formato | |
---|---|---|---|
Tumedei symmetry-2021.pdf
accesso aperto
Descrizione: Article
Tipologia:
PDF editoriale
Dimensione
2.16 MB
Formato
Adobe PDF
|
2.16 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.